Measuring points |
Pen-writing
type |
1-pen, 2-pen, 3-pen or 4-pen |
Dot-printing
type |
6 points, 12 points, 24 points |
Input signals |
Universal input |
Thermocouples |
B, R, S, K, E, J, T, N, W-WRe26, WRe5-WRe26, PtRh40-PtRh20,
NiMo-Ni, CR-AuFe, Platinel II, U, L |
Resistance
thermometers |
Pt100 (IEC751-1995), Pt100 (IEC751-1983), JPt100,Pt50, Pt-Co |
DC voltage |
±13.8mV, ±27.6mV, ±69.0mV,
±200mV, ±500mV, ±2V,±5V, ±10V,
±50V, ±50V |
DC
current |
By shunt resistor |
Contact input (Option in pen-writing type, operational
record) |
Accuracy rating |
±0.1% of measuring range ± 1 digit (DC voltage,
thermocouple, resistance thermometer inputs) |
Measuring cycle |
Pen-writing
type |
About 100ms |
Dot-printing
type |
About 5 seconds/6-point, about 10 seconds/12-point, about 20
seconds/24-point
Option … About 1 second/6-point, about 2 seconds/12-point, about 4
seconds/24-point |
Reference junction
compensation accuracy |
K, E, J, T, N, Platinel II ………………………………… Maximum
±0.5°C
R, S, W-WRe26, WRe5-WRe26, NiMo-Ni, CR-AuFe, U, L
…………………………………………………………… Maximum ±1.0°C
(The above errors are added to the accuracy ratings for an internal
reference junction compensation.) |
Burnout |
For thermocouple and resistance thermometer inputs (up-scale,
down-scale or burnout disable) |
Terminal board |
Detachable |
Printing system |
Pen-writing type |
Analog
tracing |
Disposal
cartridge pen |
Digital
printing |
Plotter
pen |
Dot-printing
type |
Wire-dotting system 6-color ribbon |
Printing colors |
Pen-writing
type |
1st-pen red, 2nd-pen green, 3rd-pen blue, 4th-pen brown |
Dot-printing
type |
1 red, 2 black, 3 blue, 4 green, 5 brown, 6 purple |
Chart |
Fan-fold type, effective width 180mm, total width 200mm,
total length 20m |
Chart speed |
Pen-writing
type |
1 to 600mm/h, 1 to 200mm/min (default - 25mm/h) |
Dot-printing
type |
1 to 1500mm/h (default - 25mm/h) |
Skip function |
No display or printings of channels |
Subtract printing |
Printing of difference of measured values of one channel and
other channel (or reference value) |
Pen-lift function |
By RECORD OFF key, all pens are lifted up simultaneously.
A lever for manually lifting up/down of all pens is provided. |
Alarm type |
Absolute value, differential, rate-of-change
(Absolute value/standby and differential/standby are additionally
available in pen-writing types. Each 2 levels) |
Alarm programming |
Individual programming each channel, Maximum 4 levels/channel
(output: option) |
Rated power voltage |
100 to 240VAC 50/60Hz, universal power supply |
Power consumption |
Pen-writing
type |
About 60VA |
Dot-printing
type |
About 45VA |
Environmental conditions |
Reference operation condition |
Ambient temperature/humidity range: 21 to 25°C/45 to
65%RH Power voltage: 100VAC Ù} 1% Attitude: Left/right 0 degree,
forward tilting 0 degree, backward tilting 0 degree Warm-up time:
30minutes or longer |
Normal
operation
condition |
Ambient temperature/humidity range: 0 to 40°C/20 to 80%RH
Power voltage: 90 to 264VAC Attitude: Left/right 0 to 10 degrees,
forward tilting 0 degree, backward tilting 0 to 30 degrees |
Mounting |
Panel mounting |
Dimensions |
Pen-writing
type |
W144 x H144 x D264mm (standard specifications) |
Dot-printing
type |
W144 x H144 x D261mm (standard specifications) |
Weight |
Pen-writing
type |
About 4.0kg (4-pen/with full options) |
Dot-printing
type |
About 3.0kg (with full options) |
Standards |
CE |
EN61326 + A1 Class A, EN61000-3-2 + A14
EN61000-3-3, EN61010-1 + A2 |
UL |
UL3111-1 (pen-writing type - approval pending) |
CSA
(C-UL) |
C22.2, No. 1010 (pen-writing type - approval pending) |
IP
|
IEC529 IP54 (front protection) |